Pulse contour analysis and transesophageal echocardiography: a comparison of measurements of cardiac output during laparoscopic colon surgery

Posted in Anesthesia, Echocardiography, Laparoscopic surgery at 2:02 by Laci

By Mario R. Concha, V Mertz, L I Cortínez, K A González, J Butte

Anesth Analg 2009;109:114-118

Pulse wave analysis (PWA) allows cardiac output (CO) measurement after calibration by transpulmonary thermodilution. A PWA system that does not require previous calibration, the FloTrac/Vigileo (FTV), has been recently developed. We compared determinations of CO made with the FTV to simultaneous measurements using transesophageal echocardiography (TEE).

Methods
Ten ASA I-II patients scheduled for laparoscopic colorectal surgery were studied. A radial 20-gauge cannula was inserted and connected to a hemodynamic monitor and a FTV system for PWA and determination of CO (COPWA). TEE CO (COTEE) was determined as previously described. Measurements were made after intubation, 5 min after establishing the lithotomy position, 5 min after establishing pneumoperitoneum, every 30 min, or each time mean arterial blood pressure decreased below basal values. Statistical analysis was made with the Bland and Altman method.

Results
Eighty-eight measurements were compared. The COTEE values ranged from 3.23 to 12 Lt/min (mean 6.21 ± 1.85). Values for COPWA ranged from 2.9 to 8.5 Lt/min (mean 4.84 ± 1.14). Bias was 1.17 and limits of agreement −2.02 and 4.37. The percentage error between all COTEE and COPWA measurements was 40% (27%-50%) mean (range).

Conclusions
During laparoscopic colon surgery, clinically important differences were observed between CO determinations made with TEE and FTV.